| What:		/sys/bus/iio/devices/iio:deviceX/calibration_auto_enable
 | |
| Date:		June 2020
 | |
| KernelVersion:	5.8
 | |
| Contact:	linux-iio@vger.kernel.org
 | |
| Description:
 | |
| 		Contaminants build-up in the measurement chamber or optical
 | |
| 		elements deterioration leads to sensor drift.
 | |
| 
 | |
| 		One can compensate for sensor drift by using automatic self
 | |
| 		calibration procedure (asc).
 | |
| 
 | |
| 		Writing 1 or 0 to this attribute will respectively activate or
 | |
| 		deactivate asc.
 | |
| 
 | |
| 		Upon reading current asc status is returned.
 | |
| 
 | |
| What:		/sys/bus/iio/devices/iio:deviceX/calibration_forced_value
 | |
| Date:		June 2020
 | |
| KernelVersion:	5.8
 | |
| Contact:	linux-iio@vger.kernel.org
 | |
| Description:
 | |
| 		Contaminants build-up in the measurement chamber or optical
 | |
| 		elements deterioration leads to sensor drift.
 | |
| 
 | |
| 		One can compensate for sensor drift by using forced
 | |
| 		recalibration (frc). This is useful in case there's known
 | |
| 		co2 reference available nearby the sensor.
 | |
| 
 | |
| 		Picking value from the range [400 1 2000] and writing it to the
 | |
| 		sensor will set frc.
 | |
| 
 | |
| 		Upon reading current frc value is returned. Note that after
 | |
| 		power cycling default value (i.e 400) is returned even though
 | |
| 		internally sensor had recalibrated itself.
 |